INSTITUTE OF APPLIED PHYSICS

List

Material Analysis Laboratory 4. Transport Measurement

Magnetic Property Measurement System

Model MPMS3
Manufacturer (Country) Quantum Design
Purchase Date 2015-12-22
Purpose of Use 시료의 자기적 특성 측정

Product Description and Specifications

  1. Temperature control-feature
  • Operating Range : 1.8 K to 400 K
  • Cooling Rate : 30 K/min (300 K to 10 K stable in 15min., typical), 10 K/min (10 K to 1.8 K stable in 5min., typical)
  • Temperature Stability : ±0.5%

 

  1. Magnetic field control-feature
  • Magnetic Field Range : -70 kOe to +70 kOe
  • Field Uniformity : 0.01 % over 4 cm
  • Field Charging Rate : 4 Oe/sec to 700 Oe/sec
  • Field Charging Resolution : 0.33 Oe
  • Remanent Field : ~5 Oe (typical) when oscillating from full field back to zero

 

  1. Magnetization measurements-feature
  • Maximum DC moment : 10 emu
  • Sensitivity : 2,500 Oe: < 1×10-8 emu (with less than 10 second averaging) > 2,500 Oe: < 8×10-8 emu (with less than 10 second averaging)
  • Variable drive amplitude : 0.1 to 8 mm (peak)

 

  1. Evercool
  • Helium Cooling Ability : 9 Liters/day
  • Liquid Helium Volume : 16 Liters
MPMS